Once removing the probe Please follow the following steps:
- Prepare 1% tergazyme solution before you remove the probe.
- After removing the probe, prevent that any organic residue can dry-out on the probe shaft.
- Flush the shaft with deionized water (milliQ water), alcohol and rinse again with deionized water. While flushing the shaft be careful so that you don't break the shaft.
- Hold the silicon shaft in 1% tergazyme solution for 30-60 min.
- Flush it again with deionized water.
- Verify that the shaft looks shiny.